1-28
Getting Started
Model 2700 Multimeter/Switch System User’s Manual
Exercise 1 — Basic DMM measurements
The exercise in
measures ACV on the 10V range and stores 15 readings in the
buffer.
Table 1-6
Exercise 1—Measure AC volts - store readings in buffer
Front panel operation
Command sequence
1
For front panel operation, proceed to step 2.
For remote programming, clear the buffer
1
:
TRAC:CLE
2
Restore defaults
2
:
Press SHIFT > press SETUP > select RESTORE: FACT.
*RST
3
Select ACV function:
Press ACV.
FUNC 'VOLT:AC'
4
Select 10V range:
Press RANGE
Δ
to display “RANGE: 10V”.
VOLT:AC:RANG 10
6
Store 15 readings in buffer
3
:
Press STORE > set for 000015 RDGS > press ENTER.
SAMP:COUN 15
READ?
7
Recall buffer readings
4
:
Press RECALL > use edit keys to display readings. Press EXIT to exit
recall mode.
CALC1:DATA?
1. To avoid problems with remote programming, it is good practice to routinely clear the buffer (
TRAC:CLE
) at the beginning
of a program that performs multiple measurements (
SAMP:COUN >1
). Restoring *RST or FACTory defaults does not
clear the buffer.
2. FACTory defaults place the instrument in a continuous measurement mode. *RST places the instrument in a non-continuous
measurement mode.
3. READ? triggers and returns 15 readings. These 15 readings are automatically stored in the buffer. See Exercise 4 and 5 for more
information on the READ? command.
4. Statistics for buffer readings are also stored in the buffer. For remote programming,
CALC1:DATA?
only returns the readings that
were stored. It does not return buffer statistics.
CALC2
commands are used to calculate and return buffer statistics (see
for details).
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......