9-12
Limits and Digital I/O
Model 2700 Multimeter/Switch System User’s Manual
8.
To retain the present pulse mode setting, press ENTER. Otherwise, press
to
move the cursor to the right, press
Δ
or
∇
key to display “NO” or “YES,” and
press ENTER.
Scanning
While limits can be configured on a per scan channel basis, the digital output configura-
tion cannot. Therefore, for all scan channels that are set to use limits, the digital output
will function according to how the Model 2700 is set up when the scan is run.
Remote programing — limits and digital output
Limits and digital output commands
The limits and digital output commands are provided in
.
Table 9-2
Limits and digital I/O commands
Commands
*
Description
Def
Ref
Limit 1 commands
CALCulate3:LIMit1:UPPer <NRf>
[, <clist>]
Set HI1 limit; <NRf> = -4294967295 to
+4294967295.
1
CALCulate3:LIMit1:LOWer <NRf>
[, <clist>]
Set LO1 limit; <NRf> = -4294967295 to
+4294967295.
-1
CALCulate3:LIMit1:STATe <b> [, clist>] Enable/disable Limit 1 test; <b> = ON or OFF.
a
CALCulate3:LIMit1:FAIL?
Query test result; 0 = pass (in), 1 = fail
(high or low).
b
CALCulate3:LIMit1:CLEar
Clear fail indication.
c
CALCulate3:LIMit1:CLEar:AUTO <b>
Enable/disable auto clear; <b> = ON or OFF.
ON
c
Limit 2 commands
CALCulate3:LIMit2:UPPer <NRf>
[, <clist>]
Set HI2 limit; <NRf> = -4294967295 to
+4294967295.
2
CALCulate3:LIMit2:LOWer <NRf>
[, <clist>]
Set LO2 limit; <NRf> = -4294967295 to
+4294967295.
-2
CALCulate3:LIMit2:STATe <b>
[, <clist>]
Enable/disable Limit 1 test; <b> = ON or OFF.
a
CALCulate3:LIMit2:FAIL?
Query test result; 0 = pass (in), 1 = fail
(high or low).
b
CALCulate3:LIMit2:CLEar
Clear fail indication.
c
CALCulate3:LIMit2:CLEar:AUTO <b>
Enable/disable auto clear; <b> = ON or OFF.
ON
c
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......