Model 2700 Multimeter/Switch System User’s Manual
Scanning
7-27
Channel setup
The <clist> parameter is used to set up scan channels. For example, the following
examples show how to set up scan channel 101:
FUNC 'VOLT', (@101)
' Set 101 for DCV.
VOLT:RANG 10, (@101)
' Set 101 for 10V range.
VOLT:DIG 4.5, (@101)
' Set 101 for 4H digit resolution.
VOLT:NPLC 3, (@101)
' Set 101 rate for 3 PLC.
NOTE
In the above command sequence, channel 101 is first set for DCV before sending
the other commands to set range, digits, and rate. If channel 101 was instead set
to a different function (i.e., RESistance), the VOLT commands to set range,
digits, and rate would generate error +700 (Invalid function in scanlist).
Buffer
For front panel scanning, the reading count specifies the number of readings to store in the
buffer. For remote scanning, the sample count specifies the number of readings to store in
the buffer.
Readings stored in the buffer by the TRAC command (or by front panel data store
operation) must be cleared before sending INITiate or READ? to take the instrument out
of idle. The following command clears the buffer:
TRACe:CLEar
' Clear buffer.
Scanning commands
Scanning commands are listed in
. Additional information on these commands
follow the table.
NOTE
Query commands and optional command words are not included in
The unabridged SCPI tables are provided in
.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......