6-2
Buffer
Model 2700 Multimeter/Switch System User’s Manual
Buffer overview
The Model 2700 has a data store (buffer) to store from 2 to 55,000 readings. The
instrument stores the readings that are displayed during the storage process. Each
timestamped reading includes the buffer location number and a timestamp.
The data store also provides statistical data on the measured readings stored in the buffer.
These include minimum, maximum, average, peak-to-peak, and standard deviation.
NOTE
When scanning, the readings are automatically stored in the buffer.
NOTE
The various instrument operations, including buffer operation, are performed on
the input signal in a sequential manner. See
, for details. It includes flowcharts showing where in the processing
sequence that buffer operations occur.
Front panel buffer
Auto clear
With buffer auto clear enabled, the buffer is cleared (readings lost) before a new storage
operation starts. The buffer can be manually cleared by setting the number of readings to
store (buffer size) to 000000.
NOTE
When editing the reading count over the front panel, press AUTO to reset the
count to 000000. You can then press <ENTER> to clear the buffer.
When buffer auto clear is disabled, the buffer is not cleared and the buffer size is set to
55000. Each subsequent storage operation appends the readings to the buffer. When the
buffer fills with 55,000 readings, the storage process stops. The 55,000 readings are
cleared before the next storage operation starts.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......