Model 2700 Multimeter/Switch System User’s Manual
Basic DMM Operation
3-47
Continuity testing
The instrument can test continuity using the 2-wire 1k
Ω
range. After selecting continuity,
you will be prompted to enter the threshold resistance level (1 to 1000
Ω
). When the
measured circuit is below the set threshold level, the instrument will beep and display the
resistance readings. When the measured circuit is above the threshold level, the unit will
stop beeping and it will display the resistance value. For instance, if the measured circuit is
out of range of the continuity function (that is, greater than 1.1KOhm), the message
“OPEN” will be displayed. If the reading is below 1100
Ω
, it will be displayed. If the
reading is 1100
Ω
or above, “OPEN” will be displayed.
NOTE
The reading rate for continuity is fixed at FAST (0.01 PLC).
Limits and digital outputs cannot be used when testing continuity with the
continuity (CONT) function. If you need to use these operations, use the
Ω
2
function to test continuity.
Connections
NOTE
When using the front panel inputs, the INPUTS switch must be in the “F” (out)
position. For switching modules, it must be in the “R” (in) position.
Front panel input
When using the front panel input terminals, connect the test leads to the INPUT HI and
LO terminals as shown in
Model 7700 switching module
Connections for the Model 7700 switching module are shown in
B. Since this
is a 2-wire ohms measurement, channels 1 through 20 can be used.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......