Model 2700 Multimeter/Switch System User’s Manual
Range, Digits, Rate, Bandwidth, and Filter
4-3
Manual ranging
To change range, press the RANGE
Δ
or
∇
key. The instrument changes one range per
key press. The selected range is displayed for one second. Note that the manual range keys
have no effect on temperature (TEMP).
If the instrument displays the “OVERFLOW” message on a particular range, select a
higher range until an on-range reading is displayed. Use the lowest range possible without
causing an overflow to assure best accuracy and resolution.
Auto ranging
To enable auto range, press the AUTO key. The AUTO annunciator turns on when auto
ranging is selected. While auto ranging is enabled, the instrument automatically selects the
best range to measure the applied signal. Auto ranging should not be used when optimum
speed is required. Note that the AUTO key has no effect on temperature (TEMP).
Up-ranging occurs at 120% of range. The Model 2700 will down-range when the reading
is <10% of nominal range.
To disable auto ranging, press AUTO. This will leave the instrument on the present range.
You can also disable auto ranging by pressing the
Δ
or
∇
key, however a range change
may occur.
Scanning
When a simple scan is configured, the present function and range setting will apply to all
channels in the scan. When an advanced scan is configured, each channel can have its own
unique range setting. Details to configure and run a scan are provided in
.
For remote programming, the <clist> parameter is used to configure channels for a scan.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......