Model 2700 Multimeter/Switch System User’s Manual
Scanning
7-21
Saving setup
Up to four instrument setups can be saved in memory using the SHIFT > SAVE menu
(SAV0, SAV1, SAV2, or SAV3). A user-saved setup can also be used as the power-on
setup. A user-saved setup can be restored from the SHIFT > SETUP menu. Details on
user-setups are covered in
.
Auto scan
When auto scan is enabled, the scan operation is saved in memory. If power to the
Model 2700 is interrupted, the scan will resume when power is restored. With auto scan
enabled, the last scan setup becomes the power-on setup. It takes precedence over the
factory, *RST, or user-saved power-on setup.
Perform the following steps to set auto scan.
1.
While in the normal measurement state, press SHIFT and then SETUP.
2.
Use the
Δ
and
∇
keys to display the auto scan (AUTOSCAN) setting; N (no) or Y
(yes).
3.
To change the setting, press
to place the cursor on the setting (N or Y) and press
Δ
or
∇
to change the setting.
4.
Press ENTER to exit from the menu structure.
NOTE
With auto scan enabled, DO NOT save the present setup as the power-on default
setup. If you do so, an interrupted scan will not resume.
If during the power-up sequence the Model 2700 detects a card ID change for
any slot, auto scan configuration will disable and an interrupted scan will not be
resumed. Error +517 occurs (cannot resume scan) to indicate that the scan has
been disabled. The instrument assumes the normal power-on setup.
The Model 7706 does not support auto scan. Trying to enable auto scan with a
Model 7706 card installed will cause error -221 (settings conflict).
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......