Model 2700 Multimeter/Switch System User’s Manual
Monitor scan (analog trigger)
A channel can be assigned as a monitor channel. When the monitor channel detects that a
reading limit has been reached, the scan will be triggered to start.
There are four reading limits that can be used to trigger the start of the scan: low limit 1
(LLIM1), high limit 1 (HLIM1), low limit 2 (LLIM2), and high limit 2 (HLIM2). The scan
will start when any enabled reading limit event is detected by the monitor channel. Details
on Limits are provided in
.
NOTE
An overflow reading (“OVRFLW” message displayed) is interpreted by the
Model 2700 as a positive reading, even if the input signal is negative. This could
inadvertently trigger a monitor scan. For example, assume the monitor channel
is monitoring a negative input signal, and the instrument is configured to trigger
a monitor scan if a positive input signal is detected. If for some reason, the
negative input signal exceeds the measurement range, the overflow reading will
be interpreted as positive and trigger the start of the scan.
Perform the following steps to run a monitor scan:
NOTE
The last enabled scan function (STEP or SCAN) will be used for the monitor
scan.
1.
Perform Step 1 and Step 2 of the
“Advanced scan setup procedure,” page 7-16
, to
set up scan channels. With the channel to be used as the monitor selected, set and
enable limits as follows. Note that you only need to set values for limits that are
going to be used.
a.
Press SHIFT and then LIMITS to access the limits menu. Note that the CHAN
annunciator flashes to indicate that the menu is being used to set up a scan
channel.
b.
Use the
,
,
Δ
, and
∇
keys to set high limit 1 (HI1) and press ENTER.
c.
Set low limit 1 (LO1) and press ENTER.
d.
Set high limit 2 (HI2) and press ENTER.
e.
Set low limit 2 (LO2) and press ENTER. The instrument returns to the scan
setup menu.
f.
Press SHIFT and then ON/OFF to display the present state of LIMITS (ON or
OFF). Again, the CHAN annunciator flashes to indicate that the menu is for a
scan channel.
g.
Press the
Δ
or
∇
key to display “LIMITS:ON” and press ENTER. The
instrument returns to the scan setup menu. Note that the HIGH and LOW
annunciators are on to indicate that limits are enabled.
After all scan channels are set up, press ENTER. The present state of IMM SCAN
is Y (yes) or N (no).
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......