15-12
SCPI Reference Tables
Model 2700 Multimeter/Switch System User’s Manual
[SENSe[1]]
:CURRent[:DC]
Path to configure DC current.
✓
:APERture <n> [, <clist>]
Set integration rate in seconds (60Hz;
1.67e-4 to 1, 50Hz; 2e-4 to 1).
(Note 2)
:APERture? [<clist>]
Query aperture integration rate.
:NPLCycles <n> [, <clist>]
Set integration rate in line cycles (60Hz;
0.01 to 60, 50Hz; 0.01 to 50).
5.0
✓
:NPLCycles? [<clist>]
Query line cycle integration rate.
✓
:RANGe
Path to set measurement range:
✓
[:UPPer] <n> [, <clist>]
Select range (0 to 3.1).
3
✓
[:UPPer]? [<clist>]
Query range.
✓
:AUTO <b> [, <clist>]
Enable or disable auto range.
ON
✓
:AUTO? [<clist>]
Query state of auto range.
✓
:DIGits <n> [, <clist>]
Specify measurement resolution (4 to 7).
7
:DIGits? [<clist>]
Query resolution.
:REFerence <n> [, <clist>]
Specify reference (-3.1 to 3.1).
0
✓
:STATe <b> [, <clist>]
Enable or disable reference.
OFF
✓
:STATe? [<clist>]
Query state of reference.
✓
:ACQuire [, <clist>]
Use input signal as reference.
:REFerence? [<clist>]
Query reference value.
✓
:AVERage
Path to configure and control filter:
:TCONtrol <name>
Select filter type: (MOVing or REPeat).
(Note 3)
:TCONtrol?
Query filter type.
:WINDow <NRf>
Set filter window in % of range
(0 to 10).
0.1
:WINDow?
Query filter window.
:COUNt <n> [, <clist>]
Specify filter count (1 to 100).
10
:COUNt? [<clist>]
Query filter count.
[:STATe] <b> [, <clist>]
Enable or disable filter.
(Note 4)
[:STATe]? [<clist>]
Query state of digital filter.
Table 15-5 (continued)
SENSe command summary
Command
Description
Default
parameter
Ref
SCPI
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......