3-42
Basic DMM Operation
Model 2700 Multimeter/Switch System User’s Manual
4-wire RTD temperature measurement configuration
The Alpha, Beta, Delta, and
Ω
at 0°C parameters for the five basic RTD types are provided
. Note that these parameters can be modified using remote programming.
The steps to configure 4-wire RTD measurements are provided in
. After
pressing SHIFT and then SENSOR, the menu starts at step 1 to select measurement units.
Each time you press ENTER to make a selection, the menu will automatically go to the
next selection. After pressing ENTER for the last step, the instrument will return to the
normal measurement state.
NOTE
As shown in
, you can select the USER sensor type, but you cannot
change the USER parameters from the front panel. The parameters for the
USER type can only be set using remote programming (see TEMPerature:FRTD
commands in
).
Table 3-5
RTD parameters
Type
Standard
Alpha
Beta
Delta
Ω
at 0°C
PT100
ITS-90
0.00385055
0.10863
1.49990
100
Ω
D100
ITS-90
0.003920
0.10630
1.49710
100
Ω
F100
ITS-90
0.003900
0.11000
1.49589
100
Ω
PT385
IPTS-68
0.003850
0.11100
1.50700
100
Ω
PT3916
IPTS-68
0.003916
0.11600
1.50594
100
Ω
Table 3-6
4-wire RTD temperature measurement configuration
Step
Menu Structure
Description
1
UNITS: C, F, or K
Select temperature measurement units
(°C, °F, or K).
2
SENS: 4W-RTD
Select the 4-wire RTD transducer.
3
TYPE: PT100, D100, F100,
PT385, PT3916, or USER
Select 4-wire RTD type.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......