A-8
Specifications
Model 2700 Multimeter/Switch System User’s Manual
Calculating dBm characteristics accuracy
As an example of how to calculate the actual reading limits for a 13dBm measurement
with a reference impedance of 50
Ω
, assume an applied signal of 0.998815V. The
relationship between voltage and dBm is as follows:
From the previous example on calculating DC characteristics accuracy, it can be shown
that a measurement of 0.998815V on the 1V range has an uncertainty of ±36.9644mV, or
0.998778V to 0.998852V, using one-year specifications.
Expressing 0.998778V as dBm:
and expressing 0.998852V as dBm:
Thus, the actual reading range is 13dBm ±0.00032dBm.
dBm and dB for other voltage inputs can be calculated in exactly the same manner using
pertinent specifications, ranges, and reference impedances.
Calculating dB characteristics accuracy
The relationship between voltage and dB is as follows:
As an example of how to calculate the actual readings limits for dB, with a user-defined
VREF of 10V, you must calculate the voltage accuracy and apply it to the above equation.
To calculate a -60dB measurement, assume 10mV RMS for a VREF of 10V. Using the
100mV range, one-year, 10Hz - 20kHz frequency band, and SLOW rate, the voltage limits
are as follows:
Accuracy =
±[(0.06% of reading) + (0.03% of range)]
±[(0.0006
×
10mV) + (0.0003
×
100mV)]
±(6µV + 30µV)
±36µV
dBm
10
V
2
IN
R
REF
⁄
1mW
-------------------------------
log
=
dBm
10
0.998778V
(
)
2
50
Ω
⁄
1mW
----------------------------------------------------
log
12.99968dBm
=
=
dBm
0.998852V
(
)
2
50
Ω
⁄
1mW
----------------------------------------------------
13.00032dBm
=
=
dB
20
V
IN
V
REF
------------
log
=
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......