9-14
Limits and Digital I/O
Model 2700 Multimeter/Switch System User’s Manual
b.
CALCulate3:LIMit1:FAIL?
CALCulate3:LIMit2:FAIL?
These commands are used to query the results of Limit 1 and Limit 2:
0 = Passing (reading within the high and low limits)
1 = Failing (reading has reached or exceeded the high or low limit)
The “1” response message does not tell you which limit (high or low) has been
reached. To determine which limit has failed, you will have to read the measure-
ment event register (
c.
CALCulate3:LIMit1:CLEar
CALCulate3:LIMit1:CLEar:AUTO <b>
CALCulate3:LIMit2:CLEar
CALCulate3:LIMit2:CLEar:AUTO <b>
These commands are used to clear the fail (“1”) indications for Limit 1 and Limit
2. If auto clear is enabled for a limit, the fail indication clears when instrument
operation enters the idle state. With auto clear disabled, the fail indication will
remain until it is cleared by the :CLEar command.
Limits and digital outputs programming example
The following command sequence configures the Model 2700 to perform Limit 1 test on a
DCV reading. If the 100mV limit is reached, digital output # 2 will be pulled low. If the
-100mV limit is reached, digital output #1 will be pulled low.
NOTE
The following example can be run from the KE2700 Instrument Driver using the
example named “Limits” in
.
*RST
' One-shot measurement mode (DCV).
CALC3:LIM1:UPP 0.1
' Set HI1 limit to 100mV.
CALC3:LIM1:LOW -0.1
' Set LO1 limit to -100mV.
CALC3:LIM1:STAT ON
' Enable Limit 1.
CALC3:OUTP:LSEN ALOW
' Set logic sense to active low.
CALC3:OUTP ON
' Enable digital outputs.
READ?
' Trigger and request reading.
CALC3:LIM1:FAIL?
' Request result of limit 1 test.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......