Model 2700 Multimeter/Switch System User’s Manual
Getting Started
1-21
Saving and restoring setups
Saving a user setup
1.
Configure Model 2700 for the desired measurement application.
2.
Press SHIFT and then SAVE to access the save setup menu.
3.
Press
to place the cursor on the present setup (SAV0, SAV1, SAV2, SAV3).
4.
Use the
Δ
or
∇
key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
Saving a power-on setup
1.
Configure Model 2700 for the desired measurement application.
2.
Press SHIFT and then SAVE to access the save setup menu.
3.
Press the
Δ
key to display the present power-on (PWR-ON) setup; FACT, *RST,
SAV0, SAV1, SAV2, or SAV3.
4.
Press
to place the cursor on the present power-on setup.
5.
Use the
Δ
or
∇
key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
Restoring a setup
1.
Press SHIFT and then SETUP to access the restore setup menu.
2.
Press
to place the cursor on the present RESTORE setup (FACT, *RST, SAV0,
SAV1, SAV2, or SAV3).
3.
Use the
Δ
or
∇
key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
NOTE
If the settings for a user setup or power-on setup do not match the switching
module types presently installed in the Model 2700, error +520 (Saved setup
scancard mismatch) occurs when the setup is recalled. The scan list will reset to
the factory defaults and all channels will open. However, the saved setup is still
retained in memory and can be restored when the matching switching module is
later installed.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......