Model 2700 Multimeter/Switch System User’s Manual
Common Commands
12-7
F *RST — reset
Return Model 2700 to RST defaults
When the *RST command is sent, Model 2700 performs the following operations:
1.
Returns Model 2700 to the RST default conditions (see “Default” column of SCPI
tables).
2.
Cancels all pending commands.
3.
Cancels response to any previously received *OPC and *OPC? commands.
NOTE
For RS-232 operation (and in some cases, GBIB operation), *OPC or *OPC?
should be used with *RST, which is a slow responding command. Details on
*OPC and *OPC? are provided in this section.
G *TRG — trigger
Send bus trigger to Model 2700
Use the *TRG command to issue a GPIB trigger to Model 2700. It has the same effect as a
group execute trigger (GET).
Use the *TRG command as an event to control operation. Model 2700 reacts to this trigger
if BUS is the programmed arm control source. The control source is programmed from the
TRIGger subsystem.
NOTE
Details on triggering are provided in
Programming example
— The following command sequence configures Model 2700 to
be controlled by bus triggers. The last line, which sends a bus trigger, triggers one
measurement. Each subsequent bus trigger will also trigger a single measurement.
NOTE
The following example can be run from the KE2700 Instrument Driver using the
.
*RST
' Restore RST defaults.
TRIG:SOUR BUS
' Select BUS control source.
TRIG:COUN INF
' Set trigger layer count to infinity.
INIT
' Take 2700 out of idle.
*TRG
' Trigger one measurement.
H *TST? — self-test query
Run self test and read result
Use this query command to perform a checksum test on ROM. The command places the
coded result (0 or 1) in the output queue. When Model 2700 is addressed to talk, the coded
result is sent from the output queue to the computer.
A returned value of zero (0) indicates that the test passed, and a value of one (1) indicates
that the test failed.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......