Model 2700 Multimeter/Switch System User’s Manual
Triggering
8-17
Trigger model operation
Once the instrument is taken out of idle, operation proceeds through the trigger model
down to the device action. In general, the device action includes a measurement and, when
scanning, closes the next channel.
Control Source
, a control source is used to hold up operation
until the programmed event occurs. The control source options are as follows:
•
IMMediate — Event detection is immediately satisfied allowing operation to
continue.
•
MANual — Event detection is satisfied by pressing the TRIG key. The Model 2700
must be in LOCAL mode for it to respond to the TRIG key. Press the LOCAL key
or send GTL over the bus to remove the instrument from the remote mode.
•
TIMer — With the timer source enabled (selected), event detection is immediately
satisfied. On the initial pass through the loop, the Timer Bypass is enabled allowing
operation to bypass the Timer and continue on to the Delay block.
On each subsequent pass through the loop, the Timer Bypass is disabled. Operation is then
delayed by the Timer or the Delay. If the user-set Timer interval is larger than the user-set
Delay, the Timer will control the length of the delay. Otherwise, the length of the delay is
controlled by the user-set Delay period.
The Timer interval can be set from 0 to 999999.999 seconds. The timer source is only
available during scan operation. The timer resets to its initial state when the instrument
goes into the normal mode of operation or into the idle state.
•
EXTernal — Event detection is satisfied when an input trigger via the TRIG LINK
connector is received by the Model 2700.
•
BUS — Event detection is satisfied when a bus trigger (GET or *TRG) is received
by the Model 2700.
Delay and Device Action
— These blocks of the trigger model operate the same for both
front panel and GPIB operation. See the front panel
, for
operating information on these trigger model blocks. Also see
, for details on Hold.
Counters
— Programmable counters are used to repeat operations within the trigger
model. For example, if performing a 10-channel scan, the sample counter would be set to
10. Operation will continue until all 10 channels are scanned and measured. If you wanted
to repeat the scan three times, you would set the trigger counter to three.
For a sample count value >1, the sample readings will automatically be stored in the
buffer. For example, with sample count set to 5, the five measured readings will be stored
in the buffer. If the trigger model is configured to repeat the sample readings (i.e. trigger
count = 2), those five new readings will overwrite the original five readings in the buffer.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......