4-8
Range, Digits, Rate, Bandwidth, and Filter
Model 2700 Multimeter/Switch System User’s Manual
Rate and bandwidth
Rate
The RATE key sets the integration time (measurement speed) of the A/D converter, the
period of time the input signal is measured (also known as aperture). The integration time
affects the amount of reading noise, as well as the ultimate reading rate of the instrument.
The integration time is specified in parameters based on a number of power line cycles
(NPLC), where 1 PLC for 60Hz is 16.67msec (1/60) and 1 PLC for 50Hz and 400Hz is
20msec (1/50).
In general, the fastest integration time (0.1 PLC from the front panel, 0.01 PLC from the
bus) results in increased reading noise and fewer usable digits, while the slowest
integration time (5 PLC from the front panel, 50 PLC from the bus) provides the best
common-mode and normal-mode rejection. In-between settings are a compromise
between speed and noise.
The Model 2700 has a parabola-like shape for its speed vs. noise characteristics and is
shown in
. The Model 2700 is optimized for the 1 PLC to 5 PLC reading rate.
At these rates (lowest noise region in graph), the Model 2700 will make corrections for its
own internal drift and still be fast enough to settle a step response <100ms.
Figure 4-1
Speed vs. noise characteristics
Lowest
Noise
Region
Voltage
Noise
166.7 s
(0.01 PLC)
16.67ms
(1 PLC)
83.33ms
(5 PLC)
Aperture Time
1s
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......