Model 2700 Multimeter/Switch System User’s Manual
Buffer
6-5
Configuring timestamp
Setting time and date
For the real-time clock, the time and date is set at the factory. However, you can check and
correct the time and date as follows:
Perform the following steps to set the time:
1.
Press SHIFT and then SETUP.
2.
Use the
Δ
and
∇
keys to display SET TIME and press ENTER. The displayed
clock will be running in the hour/minute/second AM/PM format.
3.
Use the edit keys (
,
,
Δ
, and
∇
) to set the hour, minute, and AM/PM (seconds
cannot be set), and press ENTER.
Perform the following steps to set the date:
1.
Press SHIFT and then SETUP.
2.
Use the
Δ
and
∇
keys to display SET DATE and press ENTER to display the date
in the month/day/year format.
3.
Use the edit keys (
,
,
Δ
, and
∇
) to set the date (month/day/year), and press
ENTER.
Selecting timestamp
Perform the following steps to select either the real-time clock timestamp or the relative
timestamp:
NOTE
Changing the timestamp will clear the buffer if a storage is in process. The
message “BUF CLEARED” will be displayed to indicate the buffer readings
were lost. If no storage is in process, changing the timestamp will not clear the
buffer.
1.
Press SHIFT and then SETUP.
2.
Use the
Δ
and
∇
keys to display “TSTAMP.”
3.
Press the
key to place the cursor on the presently selected timestamp (REL or
RTCL).
4.
Use the
Δ
or
∇
key to display the relative (REL) or real-time clock (RTCL), and
press ENTER.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......