Model 2700 Multimeter/Switch System User’s Manual
Basic DMM Operation
3-51
Commands
1
Description
Default Ref
PERIOD function
:PERiod:THReshold:VOLTage:RANGe
<n> [, <clist>]
Select threshold voltage range;
<n> = 0 to 1010.
10
f
:PERiod:APERture <n> [, <clist>]
Set gate time for PERIOD measurements
in secs; <n> = 0.01 to 1.0.
1.0
g
CONT function
[SENSe[1]]
Optional root command.
:CONTinuity:THReshold <NRf>
Set continuity threshold in ohms;
<NRf> = 1 to 1000.
10
SYSTem:BEEPer <b>
Enable/disable beeper; <b> = ON or OFF.
ON
Set temperature measurement units
UNIT:TEMPerature <name>
Set temperature units; <name> = C, CEL,
F, FAR, or K.
h
Trigger and retrieve readings
i
INITiate:CONTinuous <b>
Enable/disable continuous initiation;
<b> = ON or OFF.
Note 4
INITiate
Trigger one or more measurements.
[SENSe[1]]
Optional root command.
:DATA[:LATest]?
Returns the last reading string.
:DATA:FRESh?
Returns the last “fresh” reading string.
FETCh?
Return reading(s).
READ?
Trigger and return reading(s).
Channel list parameter:
<clist> = (@SCH)
where:
S = Mainframe slot number (1 or 2)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
1. The <clist> parameter is used to configure one or more channels for a scan.
2. When using multiple channel operation (ROUT:MULT command) to connect a switching module channel to the DMM for
thermocouple temperature measurements, the SIMulated reference junction will be used if the INTernal or EXTernal reference
junction is selected. The “ERR” annunciator will turn on to indicate that the integrity of the temperature reading is questionable.
3. With a Model 7700, 7706, or 7708 installed, the default sensor junction is Internal. Otherwise, the Simulated (23°C) junction is
selected.
4. The *RST default is OFF, and the SYSTem:PRESet default is ON.
Table 3-7 (continued)
Basic measurement commands
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......