Model 2700 Multimeter/Switch System User’s Manual
Basic DMM Operation
3-37
reference junction may be inconvenient but it will provide more accurate temperature
measurements (assuming the user enters a precise reference temperature).
With open thermocouple detection disabled, the Model 2700 can calculate the average
temperature of two thermocouple channels using Channel Average (see
for
details). As shown in
D, one thermocouple is connected to a primary channel
(1 through 10), and the other thermocouple is connected to its paired channel (11 through
20). Channel 1 is paired to channel 11, channel 2 is paired to channel 12, and so on. Keep
in mind that a simulated reference junction (i.e. ice bath) can instead be used for these
thermocouple temperature measurements.
Figure 3-14
Thermocouple connections
Thermocouple
Ice Bath
Model 2700
A. Simulated reference junction (front panel inputs)
H
L
CH 1-10
Model 7700
Switching
Module
H
L
CH 11-20
D. Channel average calculation, internal reference junction (Model 7700)
Thermocouple
Thermocouple
Thermocouple
C. Internal reference junction (Model 7700)
H
L
CH 1-20
Model 7700
Switching Module
Thermocouple
Ice Bath
Copper wire to thermocouple
wire connection (one of two)
B. Simulated reference junction (Model 7700)
H
L
Model 7700
Switching Module
Copper wires
CH 1-20
Copper wires
Copper wire to thermocouple
wire connection (one of two)
Input HI
Input LO
+
-
+
+
+
+
-
-
-
-
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......