10-14
Remote Operations
Model 2700 Multimeter/Switch System User’s Manual
Long-form and short-form versions
An SCPI command word can be sent in its long-form or short-form version. The command
subsystem tables in
provide the long-form version. However, the short-form
version is indicated by upper case characters. Examples:
SYSTem:PRESet
long-form
SYST:PRES
short-form
SYSTem:PRES
long-form and short-form combination
Note that each command word must be in long-form or short-form, and not something in
between. For example, SYSTe:PRESe is illegal and will generate an error. The command
will not be executed.
Short-form rules
Use the following rules to determine the short-form version of any SCPI command:
•
If the length of the command word is four letters or less, no short form version
exists. Example:
:auto = :auto
These rules apply to command words that exceed four letters:
•
If the fourth letter of the command word is a vowel (including “y”), delete it and all
the letters after it. Example
immediate = imm
•
If the fourth letter of the command word is a consonant, retain it but drop all the
letters after it. Example:
format = form
•
If the command contains a question mark (?; query) or a non-optional number
included in the command word, you must include it in the short-form version.
Example:
:delay? = :del?
•
Command words or characters that are enclosed in brackets ([ ]) are optional and
need not be included in the program message.
NOTE
For fastest response to commands, always use short forms.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......