MULTICAL® 803
124
Kamstrup A/S · Technical description · 5512-2360_A1_
GB
_02.2019
7.10
Info logger
Every time the information code changes, the below registers are logged.
Register type
Description
Date (20YY.MM.DD)
Year, month and day of logging time.
Clock (hh.mm.ss)
Time
Info
Information code on above date
E1
Heat energy
E3
Cooling energy
V1
Volume
It is possible to data read the latest 280 changes of the information code as well as the dates the changes were
made. When the information code is read from the display, the latest 50 changes including dates can be read. All of
the 280 changes can also be read by means of LogView HCW.
Info event
Every change of a parameter of the info code results in an info event if the selected country code is configured with
the parameter. It is therefore not certain that all parameters result in an info event.
An info event results in accumulation of the info event counter as well as logging in the info logger. This does not
apply as long as the calculator is in transport state or if the calculator top and base are physically separated.
Reconfiguration of active parameters of the info code will influence future info codes, whereas all logged info codes
remain as they were at the time of logging.
7.11
Config logger
Every time the configuration is changed, the below register types are logged. It is possible to data read the latest 50
changes of the config log as well as the date the changes were made. The calculator permits only 50 changes unless
you break the legal seal and carry out a total reset of the calculator, which also resets the config log.
Note: The 50
th
change of configuration must be carried out on the installation site, i.e. either via the SETUP loop or
via METERTOOL HCW.
Register type
Description
Date (20YY.MM.DD)
Year, month and date of change are saved.
Hour (hh.mm.ss)
Time is saved.
Configuration number
The new configuration number is saved.
E1, E3 and V1
Meter counts are saved right after reconfiguration.
Hour counter
Hour counter is saved.
t offset
The temperature offset value is saved.
V1 meter factor
The meter factor of V1 (imp/l or l/imp) is saved.
V1 q
p
Nominal flow q
p
is saved.
Pt type
Pt type (Pt100 or Pt500) is saved.
The calculator will always carry out a config logging if the user has had access to the SETUP loop, no matter whether
the user has changed the configuration or not.