MULTICAL®
602
94
Kamstrup
A/S
∙
Technical
Description
∙
5512
‐
931_N1_GB_11.2016
10.2.1
Data + pulse inputs (67-00-10) (PCB 5550-369)
The module has a galvanically separated data port that functions with the KMP protocol. The data output can be
used for connection of external communication units or another wired data communication which is not suitable to
perform via optical communication on the front of the meter.
See paragraph 7.3 Pulse inputs VA and VB concerning functioning of the pulse inputs.
The module comprises data connection, which can be used for external data plug, designed for use with the hand-
held terminal from Kamstrup, or as a semi-permanent PC connection.
The data connection is galvanically isolated from the optocouplers which makes it necessary to use data cable type
6699-105 or 6699-106 in order to adjust the signal to RS-232 level, which is used by PC and with the hand-held
terminal from Kamstrup. See section 11
. Data communication
for information on data strings and protocols. If the
computer does not have a COM port, a data cable with USB connection, type 6699-098, can be used.