MULTICAL®
602
Kamstrup
A/S
∙
Technical
Description
∙
5512
‐
931_N1_GB_11.2016
77
MULTICAL
®
602 and in the cable gland on the 4 wire sensor. The insulation material/cover of the cable should be
selected based on the max. temperature in the installation. PVC cables are typically used up to 80 °C and in
connection with higher temperatures silicone cables are often used.
4 wire sensor set from Kamstrup has an interchangeable sensor pocket and is available in the lengths 90, 140 and
180 mm.