Calibration and verification
12.1
High-resolution energy reading
If a need for high resolution of the energy reading arises during testing and verification it can be initialised as
follows:
-
Lift up the calculator top from the base unit and wait for the display to turn off
-
Press both push buttons simultaneously while the calculator top is placed in the base unit again and keep
pressing both push buttons until the display becomes active
-
The display now shows energy with a 0.1
Wh
resolution until one of the push buttons are activated
The display example shows 345.4
Wh
which corresponds to the energy accumulated at inlet = 43.00
C and
outlet = 40.00
C and an outlet volume of 0.1 m
3
.
The high-resolution energy reading is displayed in Wh at a volume resolution of 0.01 m³
(qp 1.5 m³/h). In connection with large meters the energy shown must be multiplied by 10 or 100.
m
3
Wh
0.001
x 0.1
0.01 x
1
0.1
x 10
1
x 100
The high-resolution energy can be used for both heat energy (E1) and for cooling energy (E3).
12.1.1
Data reading of high-resolution energy
Data reading of the register ”HighRes” is possible with ID = 155.
The read value will show correct measuring unit and value irrespective of the meter size.