Page 34 of 101
Bag size
Enter the size of the PDU transmission pocket of the server.
Validate
Click it to verify the connection between the system and the
server after all information is provided.
Steps for connecting DIOCM server are as follows.
1.
Connect the system to the network of the server with a cable.
2.
Set the network of the system and the server. Input AE title, host, IP address and the information in the
table above.
3.
Click [Validate] to verify the connection between the system and the sever. A prompt box will pop up to
indicate the verification information.
4.
If the connection is successful, you can save the patient information and cines to the server.
Ensure that the system is successfully connected to the DICOM server before use. Otherwise, the
information can't be saved to the DICOM server.
4.6 Presetting the Data Import/Export
You can import or export the data in the system by the USB storage devices or restore the factory
default setting in the Preset Data screen.
After presetting the system, click [Export Preset Data] to export the data to the USB storage device.
Click [Import Preset Data] to import the system settings data in the USB storage device to the system.
Click [Restore Factory] to restore the system settings data to the factory default setting.
You can only import the system settings data to the ultrasound system with the same type provided by the
manufacturer.
4.7 System Information
You can view the hardware version, software version, probe information and other information in the System Info
screen.