8 Parameterization
8.4.5
Menu: Input/Output
Italics = Parameter can only be viewed in the "Advanced" password level
Digital Input setup
If the filling action is stopped before reaching the configured filling quantity, the filling counter is reset to
zero. The interrupted filling is
not
continued on restarting.
Menu/parameter
Value range
Description
Display/Operator Pages/
Operator Page 1 (n)
Bargraph
Flowrate [%]
Selection of the bar graph of the displayed measured
value
You can choose between the measured values dis-
played in the value range.
Current output [mA]
Menu/parameter
Value range
Description
Input/Output
... Digital Out Mode
Selection of "... Digital Out Mode" submenu
... Logic Setup
Selection of "... Logic Setup" submenu
... Pulse Setup
Selection of "... Pulse Setup" submenu
Digital Input Setup
No Function, Totalizer Reset
(all), Flowrate to Zero,
System Zero Adjust,
Totalizer Stop (All),
Dual Range
a
, Start/Stop
Batching
a
Operating mode selection for the digital input
You can choose between six operating modes:
•
Totalizer reset for all totalizers (forward, reverse
and difference totalizer)
•
Flowrate to Zero
•
System Zero Adjust
•
Totalizer stop for all counters (forward, reverse
and difference totalizer)
•
Toggling between the measuring range 1 and 2
(Q
max
and Q
max
2)
a
•
Start / Stop the filling function (batch)
a
Default setting: Flowrate to Zero
Current Output
Selection of "Current output" submenu
a
Only available for JUMO flowTRANS MAG S02/H02
?
Summary of Contents for flowTRANS MAG H01
Page 2: ......
Page 22: ...1 Safety 22...
Page 68: ...5 Electrical connection 68...
Page 72: ...6 Digital communication 72...
Page 94: ...7 Startup 94...
Page 108: ...8 Parameterization 108...
Page 109: ...109 8 Parameterization...
Page 110: ...8 Parameterization 110...
Page 111: ...111 8 Parameterization...
Page 112: ...8 Parameterization 112...
Page 152: ...8 Parameterization 152...
Page 160: ...9 Advanced diagnostics functions 160...
Page 258: ...14 Annex 258...
Page 259: ......