10 Error messages
10.3.1
Error message during startup
No sensor detected
Once the device is switched on, the calibration data of the sensor and the transmitter settings is loaded
from the SensorMemory into the transmitter. If the communication cannot be established to the Sen-
sorMemory
1
, the adjacent message appears on the LCD display.
The device restarts after the progress bar is complete until either the communication with the Sen-
sorMemory
1
is re-established or the process is canceled by selecting "Offline".
In Offline mode, the device can be operated or parameterized, but no measurement is performed.
In Offline mode, the error message "F252.017" is set.
1
The SensorMemory is a memory integrated in the sensor.
Possible cause
Remedy
Terminals D1/D2 wired incorrectly.
Check wiring
Short-circuit or wire break of wires D1/D2
Check signal cable
Jumper SW3 not correctly connected to the back-
plane
Check jumper SW3, refer to
chapter 7.2 "Configuring the current output",
page 73.
•
off: SensorMemory provided in the sensor
(standard)
•
on: no SensorMemory in the sensor
SensorMemory
1
is defective
Contact manufacturer service
-System Startup-
NO SENSOR DETECTED
Device Resetting
Offline
Summary of Contents for flowTRANS MAG H01
Page 2: ......
Page 22: ...1 Safety 22...
Page 68: ...5 Electrical connection 68...
Page 72: ...6 Digital communication 72...
Page 94: ...7 Startup 94...
Page 108: ...8 Parameterization 108...
Page 109: ...109 8 Parameterization...
Page 110: ...8 Parameterization 110...
Page 111: ...111 8 Parameterization...
Page 112: ...8 Parameterization 112...
Page 152: ...8 Parameterization 152...
Page 160: ...9 Advanced diagnostics functions 160...
Page 258: ...14 Annex 258...
Page 259: ......