173
11 Maintenance
11
Maintenance
11.1
General information
All repair or maintenance work must only be carried out by qualified customer service personnel.
When replacing or repairing individual components, original spare parts must be used.
To install, start up and maintain the JUMO flowTRANS MAG devices, it is necessary to open the housing
of the sensor or transmitter.
Be sure to observe the following safety instructions before opening and after closing the housing for re-
starting:
DANGER – Explosion hazard!
When the case lid is open, the Ex-protection becomes void.
Before opening the housing, de-energize all of the device's connecting cables and wait for at least 20
minutes.
DANGER – Explosion hazard!
Loosening the transmitter housing screws invalidates the Ex-protection. Before re-starting, tighten all
screws on the transmitter housing.
WARNING – Personal injuries!
When the housing is open, the EMC protection is limited and the protection against contact becomes
void.
Before opening the housing switch off the voltage supply.
WARNING – Impairment of the housing protection type!
If the seal (O-ring) is not connected properly or damaged, the housing protection type is limited. Before
closing the case lid, check the seal (O-ring) for damage and replace if necessary.
When closing the case lid, make sure the seal is connected properly.
CAUTION – Damage to components!
The electronic components of the circuit board may be damaged by static electricity (observe ESD
guidelines).
Make sure that the static electricity in your body is discharged before touching electronic components.
Summary of Contents for flowTRANS MAG H01
Page 2: ......
Page 22: ...1 Safety 22...
Page 68: ...5 Electrical connection 68...
Page 72: ...6 Digital communication 72...
Page 94: ...7 Startup 94...
Page 108: ...8 Parameterization 108...
Page 109: ...109 8 Parameterization...
Page 110: ...8 Parameterization 110...
Page 111: ...111 8 Parameterization...
Page 112: ...8 Parameterization 112...
Page 152: ...8 Parameterization 152...
Page 160: ...9 Advanced diagnostics functions 160...
Page 258: ...14 Annex 258...
Page 259: ......