![JUMO flowTRANS MAG H01 Operating Manual Download Page 198](http://html1.mh-extra.com/html/jumo/flowtrans-mag-h01/flowtrans-mag-h01_operating-manual_2031178198.webp)
13 Technical data
198
13.4
Transmitter JUMO flowTRANS MAG 01
13.4.1
Electrical properties
13.4.2
Electrical isolation
Current output, digital output DO1, DO2 and digital input are galvanically isolated from the sensor input
circuit and each other. The same also applies to signal outputs of versions with PROFIBUS-PA.
13.4.3
Empty pipe detection
Requirements for the function:
•
Conductivity of the medium to be measured of ≥ 20 µS/cm
•
Signal cable length of ≤ 50 m (164 ft)
•
Sensor without preamplifier
•
Nominal width DN ≥ DN 10
13.4.4
Mechanical features
13.4.5
Temperatures
13.4.6
Protection type according to EN 60529
Voltage supply
AC 100 to 230 V (-15 %/+10 %)
AC 24 V (-30 %/+10 %)
DC 24 V (-30 %/+30 %), harmonics < 5 %
Supply frequency
47 to 64 Hz
Excitation frequency
6.25 HZ, 7.5 Hz, 12.5 Hz, 15 Hz, 25 Hz, 30 Hz
(50/60 Hz voltage supply)
Power consumption
Sensor including transmitter
AC S ≤ 20 VA (switch-on current 8.8 A at
AC 230 V)
DC P ≤ 12 W (switch-on current 5.6 A)
Electrical connection
Screw terminals (maximum 2.5 mm² – AWG 14)
Compact design
Housing
Aluminum casting, painted
Paint
Paint coat ≥ 80 µm thick, RAL 5013 (cobalt blue)
Cable fitting
Polyamide
Remote mount design
Housing
Aluminum casting, painted
Paint
Paint coat, ≥ 80 µm thick,
Middle part RAL 5013 (cobalt blue)
Front lid/rear lid RAL 5013 (cobalt blue)
Cable fitting
Polyamide
Weight
4.5 kg (9.92 lb)
Storage temperature
-40 to +70 °C (-40 to +158 °F)
Ambient temperature
-20 to +60 °C (-4 to +140 °F)
Transmitter housing
IP65, IP67 (NEMA 4X)
Summary of Contents for flowTRANS MAG H01
Page 2: ......
Page 22: ...1 Safety 22...
Page 68: ...5 Electrical connection 68...
Page 72: ...6 Digital communication 72...
Page 94: ...7 Startup 94...
Page 108: ...8 Parameterization 108...
Page 109: ...109 8 Parameterization...
Page 110: ...8 Parameterization 110...
Page 111: ...111 8 Parameterization...
Page 112: ...8 Parameterization 112...
Page 152: ...8 Parameterization 152...
Page 160: ...9 Advanced diagnostics functions 160...
Page 258: ...14 Annex 258...
Page 259: ......