9 Advanced diagnostics functions
156
9.1.8
Fingerprint
The "fingerprint" database built into the transmitter can be used to compare the values at the time of fac-
tory calibration or startup with the values currently acquired.
9.1.9
Checking the grounding
This function offers the ability to check the electrical grounding quality. Flow metering is not possible
during the check.
Conditions for using the function:
•
The measuring pipe must be completely filled.
•
There must be no flow through the sensor.
•
Maximum signal cable length in case of remote mount design type: 50 m (164 ft).
•
For this function, the conductivity of the measurement medium must be between 20 µS/
cm and 20,000 µS/cm.
Additional installation conditions:
•
There must be no pre-amplifier installed in the sensor.
9.2
Conducting the grounding check
1. Use
or
to select the "Grounding Check" item.
2. Use
to start the "Grounding Check" function.
After starting the grounding check, the frequency range up to 250 Hz is measured. The 4 strongest
frequencies from the spectrum are displayed on the right side of the display.
Menu/parameter
Value range
Description
.../Diagnostics .../Diagnostics Control /
...Grounding Check
a
Grounding Check
Starts the "Grounding Check" function
Power Spectrum
Display only
Displays the four strongest amplitudes in the power
spectrum
Amplitude 1
Display only
Amplitude 2
Display only
Amplitude 3
Display only
Amplitude 4
Display only
a
Only available for JUMO flowTRANS MAG S02/H02
Summary of Contents for flowTRANS MAG H01
Page 2: ......
Page 22: ...1 Safety 22...
Page 68: ...5 Electrical connection 68...
Page 72: ...6 Digital communication 72...
Page 94: ...7 Startup 94...
Page 108: ...8 Parameterization 108...
Page 109: ...109 8 Parameterization...
Page 110: ...8 Parameterization 110...
Page 111: ...111 8 Parameterization...
Page 112: ...8 Parameterization 112...
Page 152: ...8 Parameterization 152...
Page 160: ...9 Advanced diagnostics functions 160...
Page 258: ...14 Annex 258...
Page 259: ......