13 Technical data
186
13.2.8
Insertion length
The flanged devices comply with the insertion lengths specified according to VDI/VDE 2641, ISO 13359
or according to DVGW (work sheet W420, design WP, ISO 4064 short).
13.2.9
Signal cable length and preamplifier
For devices in the remote mount design, the electrical connection between the transmitter and sensor is
made via a signal cable.
Transmitter housing
design
Single-compartment
housing
Ex-zone 2 or outside of Ex-area
Maximum signal cable
length
a
a
At a minimum conductivity for the measurement medium of ≥ 5 µS/cm
Without preamplifier
50 m (164 ft)
With preamplifier
200 m (656 ft)
Scope of delivery
5 m (16.4 ft) included
Signal cable, part no.
00648906
Transmitter housing
design
Single-compartment
housing
Ex-zone 1
Ex-zone 2 or
Outside the Ex-area
Maximum signal cable
length
a
Without preamplifier
50 m (164 ft)
With preamplifier
-
Scope of delivery
5 m (16.4 ft) included
Signal cable, part no.
00648906
Transmitter housing
design
Dual-compartment
housing
Ex-zone 1
Maximum signal cable
length
a
Without preamplifier
10 m (164 ft)
With preamplifier
-
Scope of delivery
10 m (32.8 ft)
fixed installation
Signal cable, part no.
00648907
406018/2-0
406018/2-1
406019/2-0
406019/2-1
IMPORTANT (NOTE)!
Additional requirements for signal cable length for use of optional functions
see chapter 9 "Advanced
Summary of Contents for flowTRANS MAG H01
Page 2: ......
Page 22: ...1 Safety 22...
Page 68: ...5 Electrical connection 68...
Page 72: ...6 Digital communication 72...
Page 94: ...7 Startup 94...
Page 108: ...8 Parameterization 108...
Page 109: ...109 8 Parameterization...
Page 110: ...8 Parameterization 110...
Page 111: ...111 8 Parameterization...
Page 112: ...8 Parameterization 112...
Page 152: ...8 Parameterization 152...
Page 160: ...9 Advanced diagnostics functions 160...
Page 258: ...14 Annex 258...
Page 259: ......