7 Startup
7.5.2
Error message "Incompatible sensor"
If the transmitter is operated with a sensor of another series, the following error message appears on the
transmitter display:
A flow of ZERO is indicated in the process display. No flow measurement is performed.
1. Use
to switch to the information level.
2. Use
or
to select the "Diagnostics" submenu.
3. Use
to confirm your selection.
When trying to start up a mixed installation, the adjacent error message appears. The device cannot per-
form measurements. The display for the current flow is ZERO. The current output assumes its pre-con-
figured state (Iout for alarm).
Make sure that the sensor and the transmitter are of the same device version (e.g., sensor 406012/2-0
JUMO flowTRANS MAG S, transmitter 406018/2-0 JUMO flowTRANS MAG S).
IMPORTANT (NOTE)!
When starting up the device, make sure that the transmitter and the sensor are assigned correctly. A
mixed operation of different sensors and transmitter versions is not possible. (
see
chapter 7.1 "Checks prior to start-up", page 73)
Boiler 9
Q
0.00
l/min
Σ+
1403.09
m3
Σ-
591.74
m3
Sensor
-Operator Menu-
Diagnostics
Operator Page 1
Operator Page
2
Back
Select
Sensor
-- F
2
48.036--
Incompatible sns
Tx+snr are not
the same series
Back
Exit
Summary of Contents for flowTRANS MAG H01
Page 2: ......
Page 22: ...1 Safety 22...
Page 68: ...5 Electrical connection 68...
Page 72: ...6 Digital communication 72...
Page 94: ...7 Startup 94...
Page 108: ...8 Parameterization 108...
Page 109: ...109 8 Parameterization...
Page 110: ...8 Parameterization 110...
Page 111: ...111 8 Parameterization...
Page 112: ...8 Parameterization 112...
Page 152: ...8 Parameterization 152...
Page 160: ...9 Advanced diagnostics functions 160...
Page 258: ...14 Annex 258...
Page 259: ......