![JDS Uniphase SWS15100 User Manual Download Page 234](http://html1.mh-extra.com/html/jds-uniphase/sws15100/sws15100_user-manual_3587431234.webp)
Reference –
219
Error
Number
Error Message
Possible Cause
Solution
4120
OWB data overflow
There are overlapping
scans or a hardware error
in the receiver detector
module.
Increase the delay between
scans, using the SOM front
panel. If problem persists,
return the detector module
to JDS Uniphase for repair.
4210
Invalid Transmitter
Polarization State
(TPS)
The data collected does
not correspond to the
expected polarization.
Check that the source
optics module is set to the
same polarization state as
the SWS application
software.
Verify that the software
version number is 2.1.0 or
greater. If it is not, call the
local JDS Uniphase
representative or the service
hotline.
4220
Requested Polarization
State is missing
A polarization state is not
available.
Verify that the source optics
module is set to provide the
requested polarization state.
Check that the polarization
controller, if installed, is
connected properly. If there
is no PDL option, verify that
the SWS software is in
Insertion Loss mode (Mode
menu).
5110
Diagnostics not
available on your
system. Call your local
JDS Uniphase
representative to have
your SOM upgraded.
Disable “Diagnostics”
in the SWS15100.INI
file.
Diagnostics are not
available on the system.
Call the local JDS Uniphase
representative to have the
SOM upgraded. Disable
“Diagnostics” in the
SWS15100.INI file.
5120
Diagnostics have not
been initialized.
Perform wavelength
calibration.
Diagnostics are not
initialized.
Perform wavelength
calibration.
(table continued)