Reference –
199
Problem
Solution
14
While making PDL
measurements, the PDL
shows only noise at all
wavelengths.
If the problem persists:
•
Take another reference.
•
Acquire a new trace.
If the problem persists, contact the local JDS Uniphase
representative or the service hotline.
15
“Calibration Completed” is
not returned after the “Re-
Calibrate” button is pressed
during performance of
wavelength calibration.
Verify re-calibration:
•
Open the SWS15100.INI file in the C:\Windows or
C:\WinNT directory.
•
Verify that the wavelengths of the two peaks are
entered correctly.
•
Verify that the values of Threshold1 and Threshold2
are low enough to ensure that the peaks can be
found.
•
If the values are not low enough, edit the
SWS15100.INI file.
•
Restart the SWS software.
•
Perform the wavelength calibration again.
If the problem persists, call the local JDS Uniphase
representative or the service hotline.
Troubleshooting Flowcharts
The flowcharts in this section (Figure 105 to Figure 111) show the logical steps to use when
troubleshooting the SWS. Ensure that the system is correctly assembled and the procedures
described in the
Installation
,
Testing the SWS Components
, and
Testing the System
sections of this manual are followed. The
SWS Performance Checklist
is used to record the
results of the testing, for reporting to the service hotline or the local JDS Uniphase
representative.
The following flowcharts are listed in this section:
•
Testing the system
see the flowcharts in Figure 105 and Figure 106
•
Testing the SOM
see the flowchart in Figure 107
•
Testing the receiver
see the flowcharts in Figure 108 and Figure 109
•
Testing for system noise
see the flowchart in Figure 110
•
Testing for SOM power
see the flowchart in Figure 111