218 – Reference
Detailed Error Messages
The error messages generated by the SWS software are listed in Table 26, with actions to take
when the messages appear.
Table 26: SWS Software Error Messages
Error
Number
Error Message
Possible Cause
Solution
2
No such file or directory A file or directory required
by the program was
deleted.
Save the data and files
created. Reinstall the
software.
12
Not enough memory
The memory is corrupted.
Reboot the system.
28
No space left on device
The disk or device is full.
Delete unwanted files or try
a new device (if it is a floppy
diskette).
100
SCPI Command error
300
Device-specific error
310
System error
311
Memory error
360
Communication error
2001
Failed to save
Workspace file
The device is full.
Free up space.
4010
OWB hardware error
There is a hardware
problem in the receiver.
Check all electrical and
optical connections and try
again. If the error persists,
repair is indicated.
4020
No controller detected
The ribbon cable from the
computer to the receiver
is loose.
Check the ribbon
connection and try again.
4030
No detector detected
The ribbon cable from the
computer to the receiver
is loose.
Check the ribbon
connection and try again.
4040
OWB timeout
A timeout occurred during
data collection.
Check all optical
connections. If the problem
persists, measure the
power of the signal going to
the input of the control
module. Check that the
SOM is operating.
(table continued)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com