102 – Operating and Maintenance Instructions
Select the Band from the menubar of the SWS software.
Select either the C- or L-band from the menu bar. You can not choose to do C and L bands
simultaneously. The default settings require the use of Acetylene 12 gas cell (as fitted within the
C Band SOM) for C Band calibration, and the Acetylene 13 cell such as that within the L Band
SOM for L Band calibration. See end of this section to make changes to these defaults when
calibrating with a different gas cell (eg., Acetylene 13 for C-Band).
Figure 84. Select the Calibration option from the menubar
Enter the Calibration Dialog Box in the SWS software.
Along the menu bar of the main window of the software, select
System>Calibration
.
Figure 85. Select the Calibration option from the menubar
A window with tabs appears. Select the Wavelength Calibration tab to see the following dialog
box.