318 Chapter 16: API Command Reference
TempScan / MultiScan User's Manual
M#
- Set Measuring Mode
MultiScan/1200 only
TYPE
System
EXECUTION
Deferred
SYNTAX
M#mode
Specify measuring mode. Valid options for
mode
are:
0
- Line-cycle
integration / high-speed multi-channel mode, and
1
- Single-channel high-
speed burst mode.
DESCRIPTION
The
Set Measuring Mode
(
M#
) command selects one of the two basic internal measuring modes supported by
the MultiScan/1200 unit:
•
Line-Cycle Integration / High-Speed Multi-Channel (or Default) Mode
. In the line-cycle integration / high-
speed multi-channel mode, the MultiScan/1200 samples channel data at 1.92 kHz or once every 520.83
µ
s.
You can select the number of samples over which to average in order to accomplish noise filtering.
•
Single-Channel High-Speed Burst Mode.
In the single-channel burst mode, the MultiScan/1200 collects
data in user-defined multiples of 256 samples at a user-defined sampling frequency between 38.5 Hz and
20.0 kHz.
For more information on MultiScan/1200 measuring mode parameters, see section
Measuring Modes
(MultiScan/1200
Only)
in the chapter
System Configuration
, and see commands
Set Burst Mode Frequency
(
F#
)
and
Set Averaging Weight
(
W#
).
EXAMPLES
PRINT#1, “OUTPUT07;M#0X”
‘ Select Line Cycle Integration/High-Speed Multi-Channel Mode
PRINT#1, “OUTPUT07;W#32X”
‘ Select number of samples = 32
PRINT#1, “OUTPUT07;Y0,10,0X”
‘ Setup to read 10
Post-Trigger
scans
PRINT#1, “OUTPUT07;C1-3,1X”
‘ Enable Channels 1 - 3 for thermocouples Type J
PRINT#1, “OUTPUT07;T1,8,0,0X”
‘ Arm the unit
PRINT#1, “OUTPUT07;@X”
‘ Trigger unit and collect scans
PRINT#1, “OUTPUT07;M#1X”
‘ Select Single Channel High-Speed Burst Mode
PRINT#1, “OUTPUT07;F#10000.0X”
‘ Set sampling frequency = 10 KHz
PRINT#1, “OUTPUT07;Y0,4096,0X”
‘ Set count to take a 1.0-megabyte amount of samples
PRINT#1, “OUTPUT07;C1,26X”
‘ Select Channel 1 for reading 250 Volts AC samples
PRINT#1, “OUTPUT07;T1,8,0,0X”
‘ Arm the unit
PRINT#1, “OUTPUT07;@X”
‘ Trigger unit and collect scans
Summary of Contents for MultiScan 1200
Page 2: ...TempScan MultiScan User s Manual...
Page 4: ...TempScan MultiScan User s Manual...
Page 10: ...vi TempScan MultiScan User s Manual Notes...
Page 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Page 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Page 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Page 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Page 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Page 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Page 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Page 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Page 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Page 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Page 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Page 364: ...354 Index TempScan MultiScan User s Manual Notes...