294 Chapter 16: API Command Reference
TempScan / MultiScan User's Manual
*S
- Set Power-Up Configuration
TYPE
System
EXECUTION
Immediate
SYNTAX
*Smode
Specify power-up configuration mode. Valid options for
mode
are:
0
- Power
up under last known configuration (default).
1
- Power up under factory
default configuration.
DESCRIPTION
With the
Set Power-Up Configuration
(
*S
) command, specifying the mode to be
0
will cause the unit to load its
last known configuration when it is powered on or when the
Reset Power-On
(
*R
) command is issued. The last
known configuration will be exactly the same as the configuration it had when it was last powered down or a
*R
command was issued. Specifying this mode will insure that the unit will always come up with the configuration
which has most recently been defined. Using this mode is convenient if it is undesirable to reconfigure the unit
every time it is powered up or a
*R
command is issued.
Specifying the mode to be
1
will cause the unit to load the factory default configuration when it is powered on or
when a
*R
command is issued. Using this mode is the same as power-cycling the unit then issuing the
Restore
Factory Settings
(
*F
) command. For more information on the factory default configuration, see section
Power-
Up
Configuration
in the chapter
System Configuration
.
Note:
Issuing the
Set Power-Up Configuration
(
*S
) command
*S1
and then power-cycling the unit, or issuing
the
Reset Power-On
(
*R
) command will cause any previously defined configuration to be irrevocably lost.
However, the calibration factors for the unit and its slaves will remain intact.
EXAMPLE
PRINT#1,"OUTPUT07;Q?T?Y?X"
‘ Retrieve some current configuration settings
PRINT#1,"ENTER07"
‘ Get the settings
INPUT A$
Q1,7,7,7,0 T1,3,0,0 Y100,
1000,0
‘ Screen shows settings
PRINT#1,"OUTPUT07;*S1X"
‘ Configure to power up with factory defaults
PRINT#1, “OUTPUT07;*RX”
‘ Reset the unit
SLEEP 5
‘ Wait for the unit to reset
PRINT#1,"OUTPUT07;Q?T?Y?X"
‘ Retrieve some current configuration settings
PRINT#1,"ENTER07"
‘ Get the settings
INPUT A$
Q1,0,0,0,0 T0,0,0,0 Y0,0,0
‘ Screen shows settings
Summary of Contents for MultiScan 1200
Page 2: ...TempScan MultiScan User s Manual...
Page 4: ...TempScan MultiScan User s Manual...
Page 10: ...vi TempScan MultiScan User s Manual Notes...
Page 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Page 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Page 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Page 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Page 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Page 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Page 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Page 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Page 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Page 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Page 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Page 364: ...354 Index TempScan MultiScan User s Manual Notes...