158 Chapter 11: ScanCal
TempScan / MultiScan User's Manual
Calibration Protection
The chassis calibration constants and the calibration password are stored by the TempScan/1100 or
MultiScan/1200 in Non-Volatile RAM (NV-RAM). The password is a safety feature used to prevent
unauthorized personnel from entering
Calibration Mode
and altering the calibration constants.
As a safeguard, the calibration password and chassis calibration constants are hardware protected. This
protection is enabled by setting the microswitch 9 to the down (
0
) position on the rear panel DIP
switch. This is the default factory setting and should remain in this position unless purposely
attempting to change the password or chassis constants.
If it is necessary to change the calibration password via the
Change Calibration Keyword
(
*K
)
command, or to recalibrate the chassis, this hardware write protection can be disabled by setting
microswitch 9 to the up (
1
) position. For more information on calibration, see chapter
System
Calibration
.
CAUTION
CAUTION
CAUTION
CAUTION
To ensure accurate measurements, calibration must only be performed by
authorized personel.
CAUTION
CAUTION
CAUTION
CAUTION
Do not perform calibration until after your TempScan/1100 or MultiScan/1200
system has been powered on for at least one hour.
CAUTION
CAUTION
CAUTION
CAUTION
To ensure the integrity of the chassis calibration constants, verify that DIP
microswitch 9 is set in the down (
0
) position unless you are attempting to change
the calibration password or chassis calibration constants.
CAUTION
CAUTION
CAUTION
CAUTION
Do not forget to set DIP microswitch 9 back to the down (
0
) position when
calibration is complete. Otherwise, the calibration password and calibration
chassis constants may be corrupted and normal operation may be disrupted.
Summary of Contents for MultiScan 1200
Page 2: ...TempScan MultiScan User s Manual...
Page 4: ...TempScan MultiScan User s Manual...
Page 10: ...vi TempScan MultiScan User s Manual Notes...
Page 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Page 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Page 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Page 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Page 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Page 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Page 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Page 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Page 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Page 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Page 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Page 364: ...354 Index TempScan MultiScan User s Manual Notes...