TempScan / MultiScan User's Manual
Chapter 12: System Configuration 181
•
Auto Re-arm:
This parameter is the feature which determines whether or not, after the initial
acquisition, the acquisition will be re-enabled automatically. If the
Auto Re-arm
feature is in use,
the TempScan/1100 or MultiScan/1200 unit will immediately begin looking for another
Trigger
event once a
Stop
event and
Post-Stop
count are detected to indicate that the current acquisition
has terminated. When the next Trigger is detected, a new acquisition with the same channel
configuration will be enabled automatically. Otherwise, if this feature is not in use, and the initial
acquisition has terminated, a
Set Trigger Configuration
(
T
) command must be issued after a
Stop
event, before the unit can begin looking for a
Trigger
.
However, using
Auto Re-arm
does not allow you to change the
Trigger
configuration between
Triggers
. Consequently, the initial configuration will be used until the
Auto Re-arm
is disabled.
To disable the
Auto Re-arm
flag, a
Set Trigger Configuration
(
T
) command must be issued with
the
re-arm
parameter set to zero. Data will still be available after the
T
command is issued.
•
Trigger Synchronization:
This parameter is a flag which determines whether or not the
Trigger
(trigger start event) should be synchronized with the
Pre-Trigger
scan interval, provided that both
the
Pre-Trigger
count and
normal
scan interval are configured via the
Set Counts
(
Y
) and
Set Scan
Interval
(
I
) commands, respectively. If this flag is set, then the
Trigger
scans will begin on the
next "tick" of the
Pre-Trigger
scan interval, regardless of the exact point in time when the actual
Trigger
occurred. So, if the actual
Trigger
occurs between "ticks" of the
Pre-Trigger
scan
interval, then the acquisition will not begin until the next "tick" of the
Pre-Trigger
scan interval.
Otherwise, the acquisition of
Post-Trigger
scans will begin independently at the exact point when
the
Trigger
(trigger start event) is detected.
Trigger Start & Stop Event Sources
As already mentioned, the TempScan/1100 or MultiScan/1200 can be configured so that the
Trigger
and
Stop
are assigned different trigger sources. However, since the
Set Trigger Configuration
(
T
)
command governs the configuration of an acquisition, all other acquisition-dependent commands
should be sent before the trigger sources are assigned via the
Set Trigger Configuration
(
T
) command.
The required acquisition-dependent commands will depend on the selected trigger sources. For
instance, if the unit is configured for
Trigger On
(
@
) command, it will not be necessary to send the
unrelated
Set Trigger Level
(
L
) command.
Summary of Contents for MultiScan 1200
Page 2: ...TempScan MultiScan User s Manual...
Page 4: ...TempScan MultiScan User s Manual...
Page 10: ...vi TempScan MultiScan User s Manual Notes...
Page 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Page 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Page 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Page 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Page 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Page 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Page 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Page 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Page 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Page 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Page 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Page 364: ...354 Index TempScan MultiScan User s Manual Notes...