114 Chapter 9: TempView
TempScan / MultiScan User's Manual
•
Stop Event.
The
Stop Event
area
(b)
sets a stop scan in the acquisition by using one of the
following options (note that the
Keyboard
is not an option):
External TTL:
Use a
Rising
or
Falling
external signal.
Channel Value:
Use
Above
or
Below
channel value for a specified channel.
Alarm:
Use a specified
On
or
Off
condition.
Absolute Time:
Use a specific time of day, and date. See Notes below.
Count from Trigger:
Specify a number of scans to be read before a stop scan is set.
•
Scan Counts.
The
Scan Counts
area
(c)
specifies the number of scans – both
Pre-Trigger
and
Post
Stop
– in an acquisition.
•
Scan Intervals or High-Speed Setup.
This area
(d)
changes depending on whether the
Acquisition
Configuration
dialog box is accessed via TempScan/1100 mode, MultiScan/1200
Normal
mode, or
MultiScan/1200
High-Speed
mode.
Scan Intervals.
When in TempScan/1100 mode or MultiScan/1200
Normal
mode, the
Scan
Intervals
area appears. This area specifies the time between scans (scan frequency) – during
Pre-
Trigger
and
Post-Trigger
– and whether the same single interval is used.
High-Speed Setup (MultiScan/1200 Only).
When in MultiScan/1200
High-Speed
mode, the
High-Speed Setup
area appears. This area specifies the channel and the scan rate for high-speed,
single-channel burst mode.
•
Mode (MultiScan/1200 Only).
The
Mode
area
(e)
only appears when using a MultiScan/1200,
indicating either
Normal
or
High-Speed
mode.
Data Destination.
The
Data Destination
area
(f)
formats the data that is generated by the
acquisition. This area includes the following sub-areas:
Folder:
The
Folder
box displays the drive and directory to which data will be written.
Filename:
The
Filename
box displays the name of the data file in the above folder.
Format:
The
Format
sub-area specifies whether the data file is saved in ASCII text format
(
*.TXT
) in Microsoft Excel, or in binary format (
*.IOT
). Also, this sub-area specifies the
stamp(s) with which the data is marked –
Time/Data Stamp, Alarm Stamp,
and/or
Digital Input
Stamp
.
Auto Re-arm (New Feature):
The
Auto Re-arm
sub-area allows you to select a large number of
acquisitions to take place automatically, with each acquisition using the same configuration
settings. The following three
Capture
options are available:
•
Capture one Trigger Block
•
Capture multiple Trigger Blocks in one file
•
Capture multiple Trigger Blocks in indexed files
Summary of Contents for MultiScan 1200
Page 2: ...TempScan MultiScan User s Manual...
Page 4: ...TempScan MultiScan User s Manual...
Page 10: ...vi TempScan MultiScan User s Manual Notes...
Page 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Page 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Page 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Page 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Page 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Page 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Page 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Page 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Page 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Page 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Page 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Page 364: ...354 Index TempScan MultiScan User s Manual Notes...