TempScan / MultiScan User's Manual
Chapter 2: TempScan/1100 & MultiScan/1200 17
Calibration
Calibration of cold junction sensor:
Software control of calibrated thermocouple using the
ScanCal
program. (Calibration performed for each card and chassis in the system).
Voltage Calibration:
Software control of gain and offset.
Calibration Constants:
Chassis constants stored in NV-RAM. Card constants stored in card’s on-
board EEPROM.
Hardware Configuration
The TempScan/1100 or MultiScan/1200 unit is equipped with a high-speed IEEE 488 interface and an
RS-232/RS-422 interface. Its IEEE 488 interface is useful in laboratory applications and enables real-
time transfers of acquired data to the host computer’s hard drive for inexpensive mass storage. Its RS-
232/RS-422 interface is ideal in applications that require the placement of instrumentation at remote
distances from the controlling computer, such as process and environmental control.
This unit can be set up with an IEEE 488 or a RS-232/RS-422 interface configuration, as determined
by the DIP switch accessible from the rear panel. This DIP switch via its nine microswitches select
which command set is to be used – IEEE 488 or RS-232/RS-422 – and the operating parameters for
each. The table shows the options for its nine microswitches. Additional DIP switch settings are
shown in the following sections.
Rear Panel DIP Switch
Micro-
Label
Setting
Description
switch
#
IEEE 488
RS-232/RS-422
1
COMM SELECT
0
IEEE 488
(N/A)
1
(N/A)
RS-232/RS-422
2,3
HANDSHAKE (H/S)
00
No Handshake
No Handshake
01
Software Handshake only
(XON/XOFF) (See Note)
Software Handshake only
(XON/XOFF) (See Note)
10
Hardware Handshake only
(RTS/CTS)
Hardware Handshake only
(RTS/CTS)
11
Both Hardware and Software
Handshake
Both Hardware and Software
Handshake
4,5
IEEE ADDRESS or
00
Decimal value 0
No Parity
PARITY
01
Decimal value 8
Odd Parity
10
Decimal value 16
Even Parity
11
Decimal value 24
(N/A)
6,7,8
IEEE ADDRESS or
000
Decimal value 0
300 baud
BAUD RATE
001
Decimal value 1
600 baud
010
Decimal value 2
1200 baud
011
Decimal value 3
2400 baud
100
Decimal value 4
4800 baud
101
Decimal value 5
9600 baud (See Note)
110
Decimal value 6
(N/A)
111
Decimal value 7
(N/A)
9
CHASSIS
0
Disabled
Disabled
CALIBRATION
ENABLE
1
Enabled
Enabled
Note:
(1) XON/XOFF handshaking is valid for ASCII transmissions only. (2) At 9600 baud,
hardware (RTS/CTS) handshaking and possibly software (XON/XOFF) handshaking may be
required to maintain serial performance. However, for RS-422 operation with a Macintosh,
RTS/CTS handshaking is not recommended.
Summary of Contents for MultiScan 1200
Page 2: ...TempScan MultiScan User s Manual...
Page 4: ...TempScan MultiScan User s Manual...
Page 10: ...vi TempScan MultiScan User s Manual Notes...
Page 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Page 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Page 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Page 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Page 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Page 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Page 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Page 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Page 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Page 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Page 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Page 364: ...354 Index TempScan MultiScan User s Manual Notes...