TempScan / MultiScan User's Manual
Chapter 12: System Configuration 185
Comparing Buffered Data to Alarm Status Data
With the TempScan/1100 unit, unless scans are being collected at the maximum possible frequency, the
alarm system may detect alarm states that will not appear in the collected data. Since the alarms are
being updated at the maximum possible frequency, which may be considerably faster than the scan
frequency as set by the programmed scan interval, an alarm state may be detected even though no such
value can be found in the buffered scan data.
If such a situation is not preferred, decreasing the programmed scan interval (thus increasing the scan
frequency) will decrease the possibility of such an event. At the maximum possible scan frequency,
this situation is guaranteed not to occur.
With the MultiScan/1200 unit, the alarms are being updated at the same frequency as set by the
programmed scan interval. Hence, all detected alarm states will appear in the collected data.
Digital I/O Configuration
Located on the TempScan/1100 or MultiScan/1200 rear panel, the DB50 digital I/O connector provides
eight (8) digital input lines and thirty-two (32) digital output lines. The digital output can be controlled
either via the alarm settings or via the
Set Digital Outputs
(
O
) command. To query the digital output
state, use the
Set Digital Outputs
(
O
) command.
The
Set Digital Outputs
(
O
) command allows you to force any of the 32 digital outputs, grouped into
four 8-bit banks, to a specific setting. For each digital output, you can specify whether the bit should
be cleared with a
0
(active low, logic
false
) or set with a
1
(active high, logic
true
). This command
will override the current alarm output status as set via the
Assign Alarm Output
(
A
) command.
Each digital output line will drive five (5) standard TTL (transistor-transistor logic) loads. All digital
input lines are one-eighth (0.125) TTL loads. All inputs are protected against damage from high static
voltage. Normal precautions should be taken to limit the input voltages to the range of 0.0 to 5.3 volts.
All digital I/O lines are referenced to digital ground.
For more information on digital I/O, see section
Digital I/O Configuration
in the chapter
TempScan/1100 & MultiScan/1200
, or see command
Set Digital Outputs
(
O
) in the chapter
API
Command Reference
.
CAUTION
CAUTION
CAUTION
CAUTION
Do not exceed the levels described. Otherwise, the TempScan/1100 or
MultiScan/1200 unit may be damaged in a way that is not covered by the
warranty.
Summary of Contents for MultiScan 1200
Page 2: ...TempScan MultiScan User s Manual...
Page 4: ...TempScan MultiScan User s Manual...
Page 10: ...vi TempScan MultiScan User s Manual Notes...
Page 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Page 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Page 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Page 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Page 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Page 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Page 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Page 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Page 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Page 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Page 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Page 364: ...354 Index TempScan MultiScan User s Manual Notes...