TempScan / MultiScan User's Manual
Chapter 13: System Operation 211
Example 13b. Query & Clear Current HLL Registers (U5) with No HLL Terminator
(1)
PRINT#1, “OUTPUT07;F0,0X Q1,0,0,0,0X”
(2)
PRINT#1, “OUTPUT07;C1-4,1X”
(3)
(Command lines to configure and start an acquisition.)
(4)
PRINT#1, “OUTPUT07;U5X”
(5)
PRINT#1, “ENTER07”
(6)
INPUT A$
+1450.20 S 12:23:21.700,03/24/97
+0850.20 S 12:35:09.300,03/24/97, +0950.30
+0450.20 S 02:00:29.500,03/24/97
+0057.60 S 10:35:00.400,03/24/97, +0250.60
+0045.50 S 11:03:51.700,03/24/97
-0110.10 S 12:55:09.100,03/24/97, -0050.80
+0150.70 S 03:39:01.200,03/24/97
-0085.20 S 05:25:17.300,03/24/97, +0010.90
(7)
PRINT#1, “OUTPUT07;S?X”
(8)
PRINT#1, “ENTER07”
(9)
INPUT A$
S 13:20:00.100,03/24/97
(10)
PRINT#1, “OUTPUT07;U4X”
(11)
PRINT#1, “ENTER07”
(12)
INPUT A$
+0980.20 S 13:20:01.700,03/24/97
+0959.20S13:20:09.300,03/24/97, +0975.80
+0260.70 S 13:20:29.500,03/24/97
+0245.60S13:20:00.400,03/24/97, +0257.60
-0047.50 S 13:20:51.700,03/24/97
-0051.10S13:20:09.100,03/24/97, -0049.80
+0015.70 S 13:20:01.200,03/24/97
+0008.20S13:20:17.300,03/24/97, +0010.40
The next example demonstrates how to reset the High and Low values along with their associated
Time/Date Stamps. It also shows the effect of specifying no HLL terminator; only one
ENTER
command returns all of the HLL data.
•
Line 1:
Set the data format to Degrees Celsius (
°
C
) with HLL terminator of
None
.
•
Line 2:
Configure four channels – 1, 2, 3, and 4.
•
Line 3:
Provide the appropriate command lines to configure and start an acquisition.
•
Line 4:
After an indiscriminate amount of time, query and then reset the current HLL settings.
•
Line 5:
Get the response.
•
Line 6:
The screen will show the scan data for Channels 1 through 4 in the HLL Registers. Notice
that only one
ENTER
command returned all of the HLL data.
•
Line 7:
Now get the current time.
•
Line 8:
Get the response.
•
Line 9:
The screen will show the current time.
•
Line 10:
Now again, query (but do not reset) the HLL settings.
•
Line 11:
Get the response.
•
Line 12:
The screen will show the reset HLL data.
Summary of Contents for MultiScan 1200
Page 2: ...TempScan MultiScan User s Manual...
Page 4: ...TempScan MultiScan User s Manual...
Page 10: ...vi TempScan MultiScan User s Manual Notes...
Page 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Page 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Page 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Page 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Page 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Page 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Page 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Page 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Page 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Page 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Page 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Page 364: ...354 Index TempScan MultiScan User s Manual Notes...