1.
Remove seat. See
6.12.6 Replacing Seat Support, page
.
2.
Loosen and remove four screws (2) on the upper side of
column (3).
3.
Lift off shroud plate (1).
The shroud plate has four flaps pointing
downward. There are two tapped holes on each
flap for screws (2). By selecting upper or lower
holes, seat can be adjusted to two different
heights.
4.
Attach shroud plate (1).
5.
Install screws (2) in such a way that shroud plate is flush
fitted or is somewhat elevated depending on desired
seat height.
6.
Install parts in reverse order.
7.
Check all mobility device functions.
Fig. 4-1
4.2.2 Adjusting seat height on type A electrical lifter column with adapter
CAUTION!
Risk of crushing
The seat is very heavy. Risk of injury to hands and feet.
– Use proper lifting techniques.
CAUTION!
Risk of crushing
Risk of injury caused by uncontrolled movement of mobility device.
– Switch off power supply (ON/OFF key).
– Engage drive.
•
4 mm Allen key
•
5 mm Allen key
•
TX 40 torx wrench
•
Oblique pliers
•
Cable ties
•
Torque wrench 5 – 30 Nm (or similar)
•
Medium-strength thread-locking fluid, e.g. Loctite 243
1533060-N