With True Track Plus Motor
Fig. 6-3
Fig. 6-4
E
CLAM module
(optional)
It controls powered adjustment options. The CLAM module is mounted under the seat frame.
F
True Track Plus
power module
(optional)
It is at the rear of the battery box.
6.4 Troubleshooting
6.4.1 Operational Faults
Proceed as follows if you have any problems:
1.
First assess the possible cause of the problem using the following table.
2.
Check the remote status display. Evaluate the flash error code.
3.
Carry out the necessary checks and repairs as recommended in the following table.
The various power modules can be fitted in connection with different remotes in the mobility device. Rectification of
operational faults depends on the power module fitted. The power modules used are described in the corresponding
controls manual.
The tables for rectification of operational faults listed in the following chapters are only an excerpt from the original
manufacturer's manuals. You can obtain the original manuals from Invacare.
6.4.2 Drive fault diagnosis
Problem
Other symptoms
Possible cause
Solution
Documentation
The remote status
display illuminates
normally and shows
an error code.
Drive motors
disengaged
Engage drive motors
See corresponding
remote manual
Remote status display
does not illuminate
Batteries defective
Replace batteries
See
Completely discharged
battery
Pre-charge batteries
See user manual
Power supply to
remote interrupted
Check main fuse
See
Check cables between
modules for loose
connections or
damage
See
Remote defective
Replace remote
See corresponding
remote manual
Mobility device will
not start
Remote status display
flashing
Various causes
Assess error code
See corresponding
remote manual
1533060-N
21