1.
Turn OFF controls on remote.
2.
If necessary, remove shroud.
3.
Take note of the positions of all cables and sockets that they are connected to. Mark connectors and sockets or
take a photograph with a digital camera.
4.
Unplug actuator.
5.
Connect multimeter to the contacts and measure the resistance between the contacts. The plug can have a different
shape than shown in illustration.
A resistance below 1 ohms indicates a short-circuit. Very high resistances indicates (mega-ohms or infinity)
indicates a defective actuator. Actuator must be replaced in both cases.
5.6 Checking Battery Charge Level
The following “Dos” and “Don’ts” are provided for your convenience and safety.
DON’T
DO
Don’t perform any installation or maintenance without first
reading this manual.
Read and understand this manual and any service
information that accompanies a battery and charger before
operating the personal transporter.
Don’t perform installation or maintenance of batteries in
an area that could be damaged by battery spills.
Move the personal transporter to a work area before
cleaning terminals, or opening battery box.
Don’t make it a habit to discharge batteries to the lowest
level.
Recharge as frequently as possible to maintain a high
charge level and extend battery life.
Don’t use chargers or batteries that are not appropriate
for the chair.
Follow recommendations in this manual when selecting
a battery or charger.
Don’t put new batteries into service before charging.
Fully charge a new battery before using.
Don’t tip or tilt batteries.
Use a carrying strap to remove, move or install a battery.
Don’t tap on clamps and terminals with tools.
Push battery clamps on the terminals. Spread clamps wider
if necessary.