Invacare® Storm⁴ Series
Problem
Other symptoms
Possible cause
Solution
Documentation
Batteries defective
(unstable voltage)
Replace batteries
See
Replace motor(s)
See
and
6.6.4 Replacing
Motor-Gearbox Unit
(Storm
Mobility device
judders in drive mode
None
Drive motor(s)
defective
Replace carbon
brushes
See
or Rotating
Motor-Gearbox Unit
Sealing Ring (SSD
Motor), page 29
Mobility device pulls
to left or right
None
Drive motors running
asymmetrically
Change programming
to synchronise motors
See
Tyre visibly dented
Not enough air in tyre
Check air pressure,
replace inner tube
and/or valve if
necessary.
See
program selection
after component
replacement, page 66
Error message does
not clear
None
Bad connections
Check all connecting
cables.
See
Motor brake defective
Measure internal
resistance of brakes,
replace motor if
defective.
Motors stop and start
again
None
Voltage decline
Stop driving and allow
controls to cool down.
Motor runs but loses
power
None
High motor load
allows power module
to lower voltage
Stop driving and allow
controls to cool down.
Motors stop and do
not start again
None
High motor load
allows power module
to lower voltage
Leave mobility device
switched on and
let power module
operate. Charge
batteries overnight
with mobility device
switched on.
None
Fuse burnt out
Check cabling and
replace fuse
See
and
Motor defective
Check carbon brushes
and replace if
necessary
See
or Rotating
Motor-Gearbox Unit
Sealing Ring (SSD
Motor), page 29
None
Measure internal
resistance of motor,
replace motor if
defective.
None
Power module
defective
Replace power module
See
LiNX power module
bracket, page 60
or
True Trac Plus power
module, page 60
22
1533060-N