Component
Check
Remedy
Notes
Check contacts and
terminals
Clean contacts and
terminals
See safety information
under
and
Handling Damaged
Batteries Correctly,
page 75
Remote / Power
modules
Remote, status display
blinking
Evaluate error/blink
code
See remote manual
and controls manual.
Fixings
Tighten fixings, replace
if necessary
Cables and connecting
plugs
Tighten cables and
connecting plugs,
replace if necessary
Replace joystick knob
if necessary
Joystick function
Replace remote if
necessary
Power supply
Tighten cables and
connecting plugs,
replace if necessary
Drive program
Check Chair
configuration file
Update software
if newer version
available.
See “Updating
software”
Screws
Check screws for tight
fit
Tighten screws if
necessary
6.6 Drive Components
6.6.1 Replacing Motor-Gearbox Unit
CAUTION!
Risk of crushing
Mobility device is very heavy. Risk of injury to hands and feet.
– Use proper lifting techniques.
CAUTION!
Risk of injury caused by uncontrolled movement of mobility device
– Switch power supply off (ON/OFF key).
– Engage drive.
– Before raising mobility device, secure wheels by blocking them with wedges.
CAUTION!
Risk of crushing
The motor-gearbox unit is extremely heavy. Risk of injury to hands.
– Note high weight.
•
6 mm Allen key
•
13 mm wrench
•
Oblong wooden blocks (at least 14 x 14 x 30 cm)
1533060-N
27
Summary of Contents for Storm 4 Series
Page 4: ...Invacare Storm Series 7 Accessories 97 7 1 Accessories list 97 4 1533060 N...
Page 98: ...Notes...
Page 99: ...Notes...
Page 100: ...Notes...
Page 101: ...Notes...