62 • Chapter 4 • Preparing the CUSA
®
NXT for Surgery
Connect Tubing to Console
Follow the diagram on the side label of the Console:
1.
Route the smaller irrigation tubing through the peristaltic pump (
1
) and
secure latch.
2.
Route the aspiration tube through the pinch valve (
1
) and connect to the
patient port on the suction canister.
Warning
To avoid injury to surgical personnel, keep fingers away from the pump rollers.
Notice
When preparing the tubing set, connect the tubing to the pump prior to
attaching the irrigation supply. Attaching the irrigation supply first will fill the
tube with saline and affect the priming time.